Maintenance and cleaning
Daily maintenance:
The temperature in the vaccine compartment
must always be monitored on the thermo-
meter every morning and evening. Please
note that the vaccine in the appliance may
not freeze. The temperature is factory set,
and should not be adjusted by unauthorized
personnel.
Please check each day that the internal lid is
placed correctly and that the lid fits tightly to
the appliance.
Weekly maintenance: Remove any water at
the bottom of the refrigerator with a cloth.
Wipe of water droplets on the inside wall.
Monthly maintenance::
Clean grille for compressor compartment.
Clean the refrigerator with lukewarm water
and mild detergent.
6 Month maintenance:
Clean condenser coils.
Maintenance of the solar panels
Weekly maintenance:
The solar panels are to be cleaned for dust
once each week – or as required. Cleaning
must be carried out with water using a soft
cloth or wash rag.
Monthly maintenance:
Avoid shading. It should be regularly moni-
tored that shades do not come up such as
new towering trees, as this will decrease the
amount of energy produced by the system.
Yearly maintenance:
Electric connections and components are to
be checked and cleaned at least once a year
or more often if required. Because of risk of
fire, it is necessary to remove dust and dirt.
Checking that PV panels are clean, free of
fractures, scratches, corrosion, moisture
penetration and browning.
Cabling should be checked to ensure it is
secure.
Checking the mounting hardware to ensure
it is in good condition and ensuring the earth
connection is continuous
Checking of junction boxes to ensure there is
no water accumulation and that the integrity of
lid seals, connections and clamping devices is
intact.
Checking of junction boxes to ensure there is
no water accumulation and that the integrity of
lid seals, connections and clamping devices is
intact.
All bolts and nuts are to be re-tightened 2
weeks after installation, and then once a year.
Cleaning
WARNING:
Switch off the power supply on the front of the
unit.
The best way to clean the appliance is by
using luke warm water with a small amount
of unscented detergent. Never use cleaning
agents that scour. Use a soft cloth. Rinse with
clean water and dry thoroughly. It is important
to prevent water from running into the control
panel.
The sealing strip around the lid must be
cleaned regularly to prevent discolouration
and prolong service life. Use clean water. After
cleaning the sealing strip, check that it contin-
ues to provide a tight seal.
If the appliance is not being used for any peri-
od of time, switch the appliance off, disconnect
the power supply, empty the appliance, clean
the inside, and leave the lid open to allow air
circulation and prevent smells.