Avoid using cleaning and disinfection products
containing flammable agents (or other corrosive
agents such as acetone, chlorine or bleach).
Otherwise, ensure that the product has completely
evaporated and that there are no combustibles on the
unit and its accessories before operation.
Do not use abrasive products to clean the unit.
Do not immerse the unit.
Never use an ultrasound bath for cleaning, either the
Mini LED Black or its accessories.
None of the accessories are delivered sterilized.
Only the light guide and the flexible optical shields
(option) are sterilizable.
Before sterilization, check that your autoclave is
clean and that the water used is of good quality.
It is necessary to leave the sterilized items to cool
down to the ambient temperature and dry before re-
using them.
It is recommended to sterilize the items identified as
sterilizable by Satelec individually in the sterilization
bags foreseen for this purpose.
To maintain the sterile and aseptic state of
accessories, make sure that they are kept in
hermetically-sealed bags or containers suited to use
in dentistry.
The maintenance and/or sterilization instructions
that follow must be applied before each use of the
unit.
6.1 – Pre-disinfection/cleaning
Clean and disinfect the body, light guide and the rigid
protection shield of the Mini LED Black using ready-to-
use cleaning/disinfecting wipes based on alcohol,
amphoteric disinfectant and biguanide (refer to the
manufacturer's instructions) for at least two minutes.
Leave the product to act for at least 15 minutes.
Use wipes with CE marking, or compliant with any
standard that may be required by national regulations.
6.2 - Drying
Dry using a clean, single-use, non-woven cloth to
remove any liquid traces.
6.3 - Packing
Pack in single-use sterilization bags or sleeves compliant
with the requirements defined in the EN ISO 11607-1
standard, or compliant with any standard that may be
required by national regulations.
6.4 - Sterilization
The optic fiber light guide and the flexible optical
shields (option) for the Mini LED BLACK must be
sterilized in an autoclave according to the following
parameters:
- Autoclave: Type B compliant with the EN 13060
standard.
- Sterilization temperature: 134°C.
- Sterilization steady state: 18 minutes.
- Pressure: 2 bar minimum.
CAUTION: Not all autoclaves can reach 134°C. Not all
autoclaves perform a pre-depressurization. For further
information on the applicable sterilization instructions,
consult the autoclave manufacturer.
6.5 - Storage
After the sterilization process, store items in a dry, dust-
free place.
Before re-use, in the event that the packaging's
integrity does not conform to standard, re-package and
re-sterilize according to the defined protocol.
VII - TROUBLESHOOTING
In the event of a problem, before contacting the after-
sales service of SATELEC or the supplier:
- Make sure that the screws on the battery compartment
cover are correctly tightened.
- Check that the battery polarity is correct.
- After each use, check that there is no composite
residue adhering to the light guide. If this is the case,
immediately remove the residue and check that the
surface of the light guide has not been damaged. If
damage is visible, replace the light guide as the lamp's
power could be significantly reduced.
- Under normal conditions of use, the lamp's power does
not vary if the batteries are properly charged.
- Check that the indicator light shines green during
normal conditions of use.
- After removing the light guide, check that the LED is
clean and undamaged. If necessary, clean it using a
medical-quality dry air jet (free from compressor oil).
In the event of an anomaly, contact the supplier of
the unit rather than an unauthorized repairer, who
may return your unit in a hazardous state for both
you and your patients.
The technical service of your supplier is at your
disposal for any technical problems encountered on
your unit.