The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Magnetic field at 50Hz.
3A/m
IEC61000-4-8
Electrostatic discharge
± 6KV contact
(ESD)
± 8KV air
IEC 61000-4-2.

10.3 - Electromagnetic immunity / mobile radiofrequency equipment

The device is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the Mini LED Black (including its cables) at a distance less than
Radiated radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic environment measurement (a), must be
less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile telephones
(cellular/cordless), mobile radios, amateur radios, AM/FM radio broadcasts and TV broadcasts cannot be
determined exactly by theory. To assess the electromagnetic environment due to fixed radiofrequency emitters,
an electromagnetic environment measurement must be made. If the measured radiofrequency field strength in
the immediate environment where the product is used exceeds the compliance level specified above, the
performance of the product must be tested to verify whether it conforms to the specifications. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.

10.4 - Recommended separation distances

The device is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances are
controlled.
The device user and/or installer can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile radiofrequency communications equipment (emitters) and the device, according to the
maximum output power of the equipment, as recommended in the table below.
IEC 60601 test level
Compliance level
3A/m
± 6KV contact
± 8KV air
IEC 60601 test level
that recommended and calculated according to the frequency and power of the emitter.
3 V/m
80 MHz to 2.5 GHz
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Compliance level
d = 1.2
d = 2.3
3 V/m
Where P is the maximum power rating of the emitter in watts (W)
according to the manufacturer's specifications and d is the
recommended minimum separation distance in meters (m).
10
Electromagnetic environment - guidance
Electromagnetic environment - guidance
80MHz to 800MHz.
800MHz to 2.5GHz.