All manuals and user guides at all-guides.com
III - TROUBLESHOOTING
F
AULT FOUND
No operation
• Defective connection of the power cord
• Blown fuse(s)
• Presence of liquid or moisture between
No spray
• Defective connection of the water cord
• Insufficient water pressure
• Blocked filter or defective electro
No water at the tip but it
• Blocked tip
vibrates
• Spray not set properly
Lack of powerWeak
• Worn or misshapen tip (see TIPCARD)
vibration
• Incorrect use: wrong angle of attack or
• Presence of liquid or moisture between
No ultrasound
• Tip not tightened correctly
• Defective connector contact
• Wire in the handpiece cord cut
Water leakage between the
• Handpiece 1.15x1 watertight O-ring
handpiece and the cord, or
at the handpiece front/rear
junction
ULTRASOUND MODE
P
OSSIBLE CAUSE
the handpiece and the cord
valve
inadequate pressure on the tooth
the handpiece and the cord
worn
11
R
EMEDY
• Check the mains socket
• Return to SATELEC After
Sales Service
• Change the fuses
• Dry out any moisture
especially in the electrical
contacts
• Check the water supply
• Check the pressure
• Clean or change the filter,
otherwise return to
SATELEC After Sales
Service
• Unblock
• Adjust the spray
• Replace tip
• See tips use instructions
• Dry electrical contacts
• Re-tighten with the
wrench
• Clean the cord's contacts
• Return to SATELEC After
Sales Service for the cord
to be changed
• Change the O-ring