3-D: Precautions for Tips
For proper installation of the tips, please follow the instructions below.
A. Screw and secure the tip clockwise on the hand piece with your fingers.
B. Insert the attached tip into the torque wrench opening.
C. Ensure that the root of the tip is in the opening of the torque wrench.
D. Rotate the torque wrench clockwise until it is tight; proceed to rotate the wrench clockwise
for 3 more turns. The tip is now secured.
CAUTION: Do not 'rock' the torque wrench while installing the tip! Lateral movement
can result in a broken tip and hand piece, which is not covered under warranty. Always
be sure to thread the tip carefully prior to using the wrench to avoid cross-threading.
Section IV: Infection Control
4-A: General Infection Control Recommendations:
As with all dental procedures, use standard protective equipment such as face masks, eyewear,
face shields, gloves and protective gowns.
To ensure safety to the operator and patient, carefully follow the Infection Control Procedures
detailed in Section IV.
4-B: Water Supply Recommendations
All dental water supply systems should conform to applicable Centers for Disease Control
and Prevention (CDC) and American Dental Association (ADA) standards. These standards
should apply to flushing, chemical flushing and general infection control procedures.
4-C: Cleaning and Sterilization:
All sterilization procedures must be followed in accordance with the EN ISO17665 standards.
Autoclave sterilization conditions:
In order to avoid bacterial or viral infections, always clean and sterilize the following
components after each treatment:
1. Hand piece
2. Tips
3. Torque Wrenches
- 6 -