Les langues disponibles

Les langues disponibles

4. Seal the hand piece in a sterilization pouch (without any tips).
5. Follow the instructions of your autoclave machine to sterilize the hand piece in the
autoclave machine. The components are composed of materials that can withstand a
maximum temperature of 135℃ or (275℉) for 3 minutes. Items should take
approximately
at least 16 minutes
Preparations for Autoclave sterilization of the tip(s)
1. Clean the tip(s) preferably in an ultrasonic tank with mild disinfectant containing a
neutral pH.
2. Rinse the tip(s) in distilled water.
3. Dry the tip(s).
NOTE: Before starting the sterilization cycle, make sure that the inside of the tip is
completely dry by blowing air through the internal hole with the syringe. This will
prevent stains and patches from appearing on the surface of the tip.
4. Seal the tip inside a sterilization pouch. If you are autoclaving more than 1 tip, place each
tip in an individual bag.
5. Autoclave the tip.
Preparations for Autoclave sterilization of the torque wrench
1.
Clean the wrench.
2.
Disinfect the wrench with a cloth and mild disinfectant containing a neutral pH.
3.
Seal the wrench in a sterilization pouch.
4.
Autoclave the wrench.
Section V: Installation
5-A: General Information
If the installation of your iM3 P6 system is performed by non-iM3 distributor personnel,
check to see that requirements below are followed.
5-B: Water Line Requirements
 Incoming water supply line pressure to the iM3 P6 must be 25 psi (172 kPa minimum) to
60 psi (414Kpa maximum). If your dental water supply line pressure is above 60 psi,
install a water regulator on the water supply line to your iM3 P6 unit.
 After the above installation requirement is fulfilled, thoroughly flush the water prior to
connecting to the scaler.
to dry.
- 8 -