Press ZERO/MENU.
Display:
2Ero on
Press SET/OK to confirm.
The display and MIN/MAX values are set to zero.
or
Display:
oFL 2Ero
Cause: Pressure > 5 % of range.
Release the pressure and try again.
Resetting the zero point correction
Active: Until the device is switched off
•
Inactive: After restarting the device
•
5.2 Device Settings
1. Press and hold the ZERO/MENU button for 2 seconds.
2. Press ZERO/MENU to navigate between the options:
a) Automatic switch-off (
b) Measuring unit (
c) Display filter (
d) Delete all measurement data** (
e) Device information
3. Confirm your selection with SET/REC.
Notes:
* Values: 0-7 (0 = no damping)
** Only version with data logger
Automatic switch-off
Function activated (
The device switches off after 5 minutes of inactivity.
•
Exception: Deactivated during data recording (only version with data logger).
•
Function switched off (
Continuous operation.
•
Manual switch-off via the ON/OFF button required.
•
Remarks:
The current status (
•
start of the digital pressure gauge.
The selected setting is saved and is active again when
•
the device is switched on.
Technical changes reserved
(2 s).
(2 s).
/
Po off
Po on
)
unit
)*
Filt
dEL dAtA
):
Po on
):
Po off
/
) is displayed at the
Po on
Po off
)
)
Operation