Safety Instructions
Read the operating manual carefully. Follow all instructions to avoid personal injury and dam-
age to property.
Intended use
The digital pressure gauge may only be used for checking, adjusting and calibrating pressure
measuring devices.
WARNING
The digital pressure gauge is not a safety component in accordance with Di-
rective 2006/42/EC (Machine Directive).
 Never use the device as a safety component.
Ambient conditions
Operating temperature
Medium temperature
Storage temperature
Relative humidity
Degree of protection (EN 60529)
Vibration (IEC 60068-2-6)
Shock (IEC 60068-2-29)
WARNING
The maximum overpressure values (Pmax) specified for each pressure range
must not be exceeded. Otherwise the device may be damaged or destroyed.
Overload values
Pressure range
-1...3 bar
-1...5 bar
-1...10 bar
-1...20 bar
-1...40 bar
-1...60 bar
0...100 bar
0...160 bar
0...250 bar
0...400 bar
0...700 bar
0...1000 bar
Technical changes reserved
Overload
12 bar
18 bar
30 bar
50 bar (E2)
48 bar (D2 / C2)
80 bar
120 bar
200 bar
320 bar
500 bar
800 bar
1200 bar
1500 bar
Safety Instructions
0...50 °C
-20...+80 °C
-20...+60 °C
< 85%
IP 67
10...500 Hz / 5 g
11 ms / 25 g
Burst pressure
20 bar
30 bar
50 bar
250 bar (E2)
80 bar (D2 / C2)
400 bar
550 bar
800 bar
1000 bar
1200 bar
1700 bar
2400 bar
2500 bar