Operating messages – Error codes
GB
Hı
Lo
btLo
Store
Date
Cal
RESULT
RESULT
Time
Cal
E27 / E28 / E29
E 10 / E 11
E 20 / E 21
E23 / E24 / E25
E 22
E 70
E 71
E 72
E 73
Copyright and Trademark Notice
The Bluetooth
word mark is a registered trademark owned by Bluetooth SIG, Inc. and any
®
use by The Tintometer
iOS
is a registered trademark of Cisco, Inc. and licensed to Apple, Inc.
®
iTunes Store
is a trademark of Apple, Inc., registered in the U.S. and other countries.
®
Android
and Google Play
™
Excel
is a trademark of Microsoft Corp., registered in the U.S. and other countries.
®
MD110_3b 12/2020
Operating messages
Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
A user calibrated method is indicated by an arrow while the test result
is displayed (see "Factory calibration reset").
Error codes
Light absorption too great. Reasons: e.g. dirty optics.
Calibration factor "out of range"
Too much light reaching the detector.
Too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.
LR:
Factory calibration incorrect / erased
LR:
User calibration incorrect / erased
MR:
Factory calibration incorrect / erased
MR:
User calibration incorrect / erased