A p p L I C AT I o N
8.4.2. MOuNTINg THE ILA/I-COr MEMBrANE LuNg
risk of serious harm to patient.
WArNINg
Damage of membrane lung caused by use of tools for
fixation could result in serious harm to patient
Do not use tools to tighten securing screw of bracket
1
Turn the screw to open the bracket until the membrane lung can be
inserted
2
To fix the membrane lung onto the bracket close the screw [fig
3
Ensure that the bracket engages the ribs of the membrane lung
10
[fig
]
8.5. FASTENINg THE PuMP DrIVE
1
Attach the pump drive to the top of the compact holder Make sure
that the pump drive is fully seated in the compact holder
2
Tighten the clamping screw by hand [fig
8.6. ADJuSTINg THE PuMP DrIVE INCLINATION ANgLE
The angle of inclination of the pump drive can be adjusted
1
To do this, open the clamping screw and bring the pump drive into the
desired position
2
Fix the desired position by tightening the clamping screw [fig
8.7. ATTACHINg THE SENSOr BOX
Attach the sensor box to the outer right side of the compact holder
1
To do this, use the fastener on the back of the sensor box
2
Ensure that the sensor box is attached in a way that it cannot slip off
the side of the compact holder