D A N G E R S , WA R N I N G S A N D C A U T I o N S
Danger to life of patient
Use of xenios system by unqualified staff will result in death
or serious injury of patient only specially trained, qualified
medical professionals are allowed to use xenios system
DANgEr
Do not use system without training for xenios system and
treatment mode in use
Danger to life of patient
Do not modify the xenios system
Do not install or use unauthorized software
Do not modify holding system
risk of critical harm to patient
obstruction including kinking of blood lines can lead to
diminished blood flow and could result in critical harm to
patient
Set flow alarm and lower flow limits
Closely monitor pressures within system and set alarms
accordingly
Attach flow/bubble sensor only to marked indicator on
corresponding tubing kit
Regularly check tubing for kinks Attach components in a
WArNINg
way that prevents kinking of the tubing
risk of serious harm to patient
If holding system was dropped or fell, all components must
be checked for damage, including holding system and all
fastening components
Do not use xenios system if it is damaged, then device
must be exchanged
risk of critical harm to patient
position tubing set as low as possible
Ensure pump drive is slightly below or at least at same
height as oxygenator
EN
13