3 Ethernet (optional)
The Ethernet allows you to transmit data via cable to devices (such as computers,
printers etc.) that are interconnected in a local network. No direct connection between
KIB-TM and PC is necessary.
Set the following menu items in KIB-TM (See chap. 8)
 Menu item "P9 Prt"  "oPt"  "intF"  "EnEt" (Enable output Ethernet)
 Menu item "P9 Prt"  "oPt"  "ModE"  " Count" (Output mode cont. data
output)
 Menu item "P9Prt"  "oPt"  "iP1-4" Set IP address KIB-TM as follows:
Enter IP address not yet allocated in network:
Example: 10.0.1.104
It is always necessary to enter three numbers following scheme below:
10.
0.
1
010
000
001
IP1
IP2
IP3
The same principle is used to configure the following settings:
 Menu item "P9 Prt"  "oPt"  "MASK_1-4" (Subnet mask)
 Menu item "P9 Prt"  "oPt"  "GATE_1-4" (Gateway)
Now enter the IP address for the PC on the display unit
(If unknown proceed as follows:
 Press Windows key and "R" simultaneously
 Enter "cmd" and press Enter to confirm
 The entry prompt will appear
 Enter "ipconfig" and press Enter to confirm
 The PC's IP address will appear on the screen)