settings menu AT. Press or hold the ON/OFF button to
perform corresponding operations. Hold it to set an item, or
press it to switch an option or switch the display mode.
Press means no more than 0.5 seconds, while hold means
more than 0.5 seconds.
Alert Sound Setting
Hold ON/OFF button 1 while the device is switched
on.Settings menu (interface 1) is displayed AT-a. Move «*»
to the corresponding option, and hold the function button to
set Alert to on and set Beep to off. When Alert is set to on
and the measured values of the blood oxygen saturation
and pulse rate go beyond the upper limit or lower limit, the
device gives off an alert sound. When Beep is set to on, a
tick will be heard along with pulse beats during pulse rate
measurement. While the «*» symbol stays on the Restore
option, hold the functional button to restore default settings.
Brightness Setting
Press ON/OFF button 1 to select the Brightness option
and then hold the ON/OFF button to set the brightness to a
value ranging from 1 to 5. The greater the value, the
greater the brightness of the screen.
Alert Range Setting
On settings menu (interface 2) AT-b, press the ON/OFF
button 1 to switch between options. On this interface, you
can set the upper limit and lower limit of SpO2 Alert and
Microlife OXY 210
PR Alert. While the «*» symbol stays on the +/- option, hold
the functional button to set the option to + or -.
In + mode, select the corresponding option and hold the
ON/OFF button to increase the upper or lower limit; in -
mode, hold the ON/OFF button to decrease the upper or
lower limit. Move «*» to the Exit option, and hold the ON/
OFF button to return to the monitoring interface.
7. Display Mode
When the device is switched on, shortly press the ON/OFF
button 1 to switch to another display mode to select your
desired display mode 9. There are 6 different display
modes. The default setting is mode 1.
8. Using the lanyard
1. Thread the thinner end of the lanyard through the
hanging hole at the rear end of the device.
2. Thread the thicker end of the lanyard through the
threaded end before pulling it tightly.
9. Malfunctions and actions to take
Description
Symptom/
Possible causes
SpO
or pulse
1. Finger is not
2
inserted
rate do not display
correctly.
normally.
2. Patient SpO
value is too
low to be
measured.
3. There is
excessive illu-
mination.
SpO
or pulse
1. Finger might
2
not be inserted
rate is shown
deep enough.
unstable.
2. Excessive
patient move-
ment.
Solutions
1. Retry
inserting the
finger.
2. & 3.
2
Measure
more times.
If you deter-
mine the
product is
working
correctly,
consult your
doctor.
1. Retry
inserting the
finger.
2. Sit calmly
and retry.
5
EN