WMM30 Product HANDBOOK
Special Instructions for medical applications (IEC/EN/UL/CSA 60601-1)
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with
oxygen, or with nitrous oxide. They are NOT protected against the ingress of water.
Where any part of this product is made accessible to the operator in the end use equipment, the operator
must not touch this part and the patient at the same time. If the equipment is used in a manner not specified
by the manufacturer, the protection provided by the equipment may be impaired.
WARNING: No modification of this product is allowed.
Environmental Specifications:
Description
Use
Temperature
Humidity
Altitude
Material Group
Pollution Degree
Overvoltage Category
Safety Class
Level of Insulation:
WMM30
Primary mains circuits to secondary circuits
The WMM30 is a Class II product (no Earth)
WEEE
Reference should be made to local regulations concerning the disposal of these products at the end of their
useful life.
TDK Lambda UK Ltd is a member of the B2B Compliance scheme; the scheme is operated by GAMBICA
B2B Compliance Ltd
For further information relating to the obligations of TDK-Lambda under the WEEE Directive please contact
B2B Compliance on +44 (0)1691 676124, you can also contact them via e-mail at
WWW.b2bcompliance.org.uk
Operation
Indoor
-25°C to +70°C (derate linearly for 100% to 50%
load from 40 to 70°C)
5 to 93% RH, non-condensing
Page 13 of 16
5000m
IIIb
2
II
II
Isolation voltage
4kVac
Storage
-
-25°C to +85°C
5 to 93% RH, non-condensing
Means of protection
2 x MoPP
260676 issue 1, June 2020