CUS400M Handbook
Special Instructions for medical applications of the following standards:
CSA-C22.2 No. 601.1-M90
rd
IEC/EN/UL 60601-1 3
ANSI/AAMI ES 60601-1
CSA 22.2 No 60601-1
These products are designed for continuous operation within an overall enclosure, and must be
mounted such that access to the mains terminals is restricted. See the appropriate standard listed
above.
These products are NOT suitable for use in the presence of flammable anesthetic mixtures with air or
with oxygen, or with nitrous oxide.
rd
For IEC/EN 60601-1 3
provide reinforced insulation between input and outputs of 2 MOPPs. 1 MOPP from input to
Earth/chassis/customer fixing point and 1 MOPP from output to Earth/chassis/customer fixing point.
These products are suitable for B and BF type medical equipment.
These products are NOT protected against the ingress of water.
All outputs have basic spacing to earth rated for mains at 250Vac, and due consideration must be
given to this in the end product design.
These products have SELV or ES1 outputs.
Reference should be made to local regulations concerning the disposal of these products at the end of
their useful life.
Where any part of this product is made accessible to the operator in the end use equipment, the
operator must not touch this part and the patient at the same time.
The chassis/cover/customer fixing points when fitted to the CUS400M in a Class II configuration has
(Basic/Supplementary) insulation to Primary and (Basic/Supplementary) insulation to Secondary.
For Single fuse (/E option) the end equipment manufacturer must use a polarised input connector.

Environmental Specifications:

Description
Use
Temperature
Humidity
Altitude
Pressure
Orientation
Material Group
Pollution Degree
Overvoltage Category
Class
Weight
IP Rating
Edition
Edition, ANSI/AAMI ES 60601-1, CSA 22.2 No 60601-1, these products
Operation
Indoor
-20°C to 70°C (derated linearly
from 50°C by 2.5% per °C)
5 to 95% (15 to 90% for fan versions) RH, non-
condensing
-200m to 5000m
70kPa to 106kPa
Any excluding inverted.
IIIb
2
II
I
Typical 432g (with potted baseplate)
IPX0
Page 13 of 22
Storage
-
-40°C to 85°C (70°C for /F
option)
5 - 95% (15 to 90% for fan
versions) RH, non-condensing
-200m to 5000m
54kPa to 106kPa
All
260519-02 Apr 2020