Special Instructions for medical applications of the following standards:
nd
IEC/EN 60601-1 2
Edition
UL 60601-1 1st Edition
CSA-C22.2 No. 601.1-M90
rd
IEC/EN 60601-1 3
Edition
ANSI/AAMI ES 60601-1
CSA 22.2 No 60601-1
These products are designed for continuous operation within an overall enclosure, and must be
mounted such that access to the mains terminals is restricted. See the appropriate standard listed
above.
For Class II installation, these products need to be fixed such that they are isolated from unearthed
accessible conductive parts by at least 2 MOPP's.
These products are NOT suitable for use in the presence of flammable anesthetic mixtures with air or
with oxygen, or with nitrous oxide.
nd
For IEC/EN 60601-1 2
have a reinforced insulation barrier between input and output.
rd
For IEC/EN 60601-1 3
provide reinforced insulation between input and outputs of 2 MOPPs. 1 MOPP from input to earth and
1 MOPP from output to earth.
These products are suitable for B and BF type medical equipment.
These products are NOT protected against the ingress of water.
All outputs have basic spacing's to earth rated for mains - 250Vac, and due consideration must be
given to this in the end product design.
These products have SELV outputs.
Reference should be made to local regulations concerning the disposal of these products at the end of
their useful life.
Where any part of this product is made accessible to the operator in the end use equipment, the
operator must not touch this part and the patient at the same time.
WARNING: No modification of this product is allowed.
Special Instructions for IEC/EN/UL/CSA 61010-1
If the equipment is used in a manner not specified by the manufacturer, the protection provided by
the equipment may be impaired.
Special Instructions for IEC/EN 60335
For IEC/EN60335 applications, it must be ensured within the end installation, the earth terminal is
marked with a protective earth symbol 5019 of IEC60417.

Environmental Specifications:

Description
Use
Temperature (1)
Humidity
Altitude
Pressure
Orientation
Material Group
Pollution Degree
Overvoltage Category
Description
Class
Weight
IP Rating
* - 3000m for 60601-1
(1) - Ripple and noise specifications applies 0 to 70°C. Start up at -20°C
Edition, UL 60601-1 1st Edition, CSA-C22.2 No. 601.1-M90, these products
Edition, ANSI/AAMI ES 60601-1, CSA 22.2 No 60601-1, these products
Operation
Indoor
-20°C - +70°C (See o/p tables for de-ratings)
5 - 95% RH, non-condensing
-200m - 5000m*
70kPa - 106kPa
Any excluding inverted.
Operation
Page 13 of 22
-40°C - +85°C
5 - 95% RH, non-condensing
-200m - 5000m
54kPa - 106kPa
IIIb
2
II
I and II
<200g
IPX0
Storage
-
All
Storage
260153 issue10, October 2018