MFC2046 Installation and Safety Manual
2-2-1

FCA7046 Amplifier Box

Environment
Operating temperature
Storage / transport
temperature
Altitude
Relative humidity
Pollution
Magnetic environment
Electromagnetic
environment
2-2-1

MFC9046 Probe-Array

Frequency
Environment
Operating temperature
Storage / transport
temperature
Altitude
Relative humidity
Pollution
Electromagnetic
environment
v 1.0 r 1.2 – 12/2018
Indoor use; IP 40
10 – 40 °C
-25 – 80 °C
≤ 2000 m
Maximum 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
Pollution degree 2: only non-conductive pollution
occurs, except that occasionally a temporary
conductivity caused by condensation is to be
expected.
< 0.2 T
Equipment intended to be used in an industrial
electromagnetic environment, class A.
See warning below.
Depends on model. See Probe-Array specification.
Indoor use
10 – 40 °C
Deuterium probe: 0 – 80 °C
Other: -25 – 80 °C
≤ 2000 m
Maximum 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
Pollution degree 2: only non-conductive pollution
occurs, except that occasionally a temporary
conductivity caused by condensation is to be
expected.
Equipment intended to be used in an industrial
electromagnetic environment, class A.
See warning below.
5
Copyright © 2018 Metrolab Technology SA