MFC2046 Installation and Safety Manual
• Use the MFCTool Instrument Connection Settings to reconnect the
instrument.
• Restart the PT2026 by powering down and back up.
• Use the MFCTool software settings to restore the factory defaults.
• Keep the power switch depressed while powering down the PT2026. After
approximately five seconds, the instrument should restore the factory
settings and then power down.
• The power-down reset described above will not work if the instrument
firmware has crashed. In this case, continue to hold the power switch, and
after approximately ten seconds, the PT2026 will be forced to power down.
• If the PT2026 firmware detects a serious fault during startup, the
instrument will reboot in Device Firmware Upgrade (DFU) mode, ready to
receive a firmware upgrade. See Section 6-Maintenance and Service.