This refrigerator has been designed to provide both an audible and visual alarm should any of
the following situations occur:
• Door is left open for a set period
• Internal air temperature goes outside the set parameters, either too high or too low
• Internal 'load' temperature goes outside the set parameters, either too high or too low
Should one of these alarms occur this can be muted by pressing
The 'load' temperature is a simulated temperature which is intended to provide an indication
of the temperature of the refrigerator's contents (e.g. a 5ml sample). The temperature of a
refrigerator's contents will change more slowly (up or down) than the air temperature within
that refrigerator should there be a fault or the door is left open.
All ES series laboratory refrigerators include remote alarm contacts and these terminal
contacts enable the user to connect to a remote control station; these contacts can be found on
the rear of the cabinet. WARNING: These contacts must only be accessed by a competent
electrician.
In order to use these contacts the following must be carried out:
1. The refrigerator MUST be disconnected from the power supply before removing the
terminal cover.
2. Unscrew the cover (figure 1) on the right hand side of the rear of the cabinet to expose
the connectors (figure 2).
3. Use the right hand outer positions that correspond to the left hand cable positions for
your connections using 0.75mm cable stripped to 0.6mm.
4. Depress the larger slot above the cable entrance on the right hand side with a small
electrical screwdriver. This allows the cable to be inserted into the slot directly below.
In the event of a door open alarm, high or low temperature alarm or power failure a remote
relay will switch within the controller to make an electrical circuit.
Warning Alarms
Remote alarm contacts
Figure 1
on the controller.
Figure 2