Figure 6 I/O module-powered loop
1 Analog output
Figure 7 Externally-powered loop
1 Analog output
2 External power supply
4.3.6 Connect the digital outputs
Do not use the digital outputs for process control functions such as pump on/off switching or water system
shutdown. The digital output connections do not replace a PLC (programmable logic controller).
The I/O module has four digital outputs. Connect a compatible device such as a remote alarm
indicator, buzzer or PLC to the digital outputs. Refer to
Table 3
to connect the digital outputs. Refer to
specifications.
The digital outputs are independent floating switches when the PULLUP V input and associated
internal pull-up resistors are not used. The switches are normally open. The switches close when the
user-selected alarm conditions occur.
To use the optional PULLUP V input and the related internal 10 kΩ internal pull-up resistors, supply
the applicable voltage to the PULLUP V and COM (pin 14) pins. Then, connect a jumper from the
OUT # LO pin to the COM pin. The output signal logic is active low.
Pin
OUT # HI
OUT # LO
COM
14 English
2 External device
3 External device
N O T I C E
Figure 4
Specifications
Table 3 Wiring—digital outputs
on page 11,
Figure 5
on page 3 for the digital output
Description
Positive
Negative
Common
on page 12 and