Figure 5 Wire installation
4.3.3 Connect the relays
Potential Electrocution Hazard. Power and relay terminals are designed for only single wire
termination. Do not use more than one wire in each terminal.
Potential fire hazard. Do not daisy-chain the common relay connections or jumper wire from the mains
power connection inside the instrument.
Fire hazard. Relay loads must be resistive. Always limit current to the relays with an external fuse or
breaker. Obey the relay ratings in the Specifications section.
The I/O module has four non-powered alarm relays. Each relay changes state when the selected
alarm condition for the relay occurs. The alarm condition includes one or more alarms in a boolean
OR or AND logic configuration. Refer to the FL1500 flow logger or AS950 controller documentation to
configure the relays.
Connect the relays to a control device or an alarm device. Refer to
on page 12 and
Table 1
specifications. The relays are isolated from each other and the low-voltage input/output circuitry.
The relay terminals accept 20 to 14 AWG wire (as determined by load application). Use wire with an
insulation rating of 300 VAC or higher.
The current to the relay contacts must be 0.5 A or less. Make sure to have a second switch available
to remove power from the relays locally in case of an emergency or for maintenance.
12 English
W A R N I N G
W A R N I N G
C A U T I O N
to connect the relays. Refer to
Figure 4
on page 11,
Specifications
on page 3 for the relay
Figure 5