Symptoms
The unloaded device does not always
come down.
The moving parts of the device are hard, the
device is hard to handle.
The cylinder does not work but there is a
«click» in the housing
control when the button is pressed.
remote control.
The actuator does not work but there is no
«click» in the control box when the remote
control is pressed.
The cylinder stops.
The cylinder no longer moves up or down.
Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)
Troubleshooting Guide
Our aircraft need to have weight on them in
order to descend.
This is due to a lack of
lubrication of moving parts.
1. The batteries are discharged.
2. The cylinder cable is not connected.
3. The cylinder cable is damaged.
4. The cylinder or control box is damaged.
1. The emergency stop button is pressed.
2. The battery is discharged.
3. The remote control cable is not connec-
4. The remote control cable is damaged.
5. The entire electrical system is damaged.
1. Battery is discharged (sound system plus
indication «low battery» on the display).
2. The appliance has been used for too long.
3. The weight lifted by the device is too
1. The control box is out of order.
2. The remote control is out of order.
Assembly Guide / Lifts : NAUSIFLY 3
Causes
ted.
great.
Solutions
Press the lift arm while pressing the down
button on the remote control.
Lubricate moving parts.
1.Load the device.
2. Connect the cylinder cable.
3. Have the cylinder repaired.
4.Have the electrical assembly serviced.
1. Turn the knob to unlock it.
2. Check the condition of the charger and
then charge the battery.
3. Connect the remote control.
4. Change the remote control.
5. Have the electrical assembly serviced.
1. Check the condition of the charger and
then charge the battery.
2. Let the appliance stand for a while (pro-
tection for the life of the cylinder).
3. Reduce the weight.
1. Have the control box serviced.
2. Change the remote control.
57