Calibrating
AutoCal Calibration
(Calibration with automatic buffer recognition)
The calibration method is selected in the configuration menu.
Calibration is required to adjust the sensor
indispensable for achieving comparable and reproducible
measurement
Measurement
pH, mV value,
temperature
mV value blinks until calibration is completed, then successive display of:
Then the meter switches to measuring mode.
Note: To abort calibration, you can press meas at any time. This will be con-
firmed by the display message "CAL ABORTED". Exception: When you have
selected "CAL POINTS 1-2-3" and the first calibration step has been complet-
ed, the calibration process cannot be stopped any more.
42
results.
Calibration method, the number of calibration points
and the buffer set have been selected in the configu-
ration menu.
Dip sensor in 1st/2nd/3rd buffer solution.
It does not matter which buffer solution is taken first.
Depending on the number of calibration points, the
procedure described for CAL 1/2/3 is repeated.
to the meter. It is