Start-Up

Connecting a Sensor

The HandyLab 750/750EX provides several connections so that many types
of sensors can be used for measurement.
Note that only one sensor may be connected to the meter at a time.
The meter automatically recognizes a connected Memosens sensor and
switches accordingly. Memosens is signaled in the display.
Caution!
Always make sure that a sensor is connected to the meter be-
fore starting measurement.
Explanation: The analog pH input of the HandyLab is an elec-
trometer amplifier with an extremely high-impedance. When the
sensor is not in contact with the medium or not connected to
the meter, electric charges on the input can generate arbitrary,
stable pH or mV values which will be shown in the display.
Connections
a
Micro USB port
b
M8, 4 pins for Memosens sensors
c
Temperature probe GND
d
Temperature probe
e
pH socket (DIN 19 262)
Memosens sensors have a cable coupling, which allows convenient replace-
ment of sensors while the cable remains connected to the meter. The con-
necting cable is connected to socket b (M8, 4 pins for Memosens sensors).
38
a
b
c
d
e