Standard Features
• Simple to operate.
• Portable and lightweight.
• Pad consists of 130 bubble cells.
Operating Instructions
1. Position pump on horizontal surface or hang it by means of the hook.
2. Open mattress on the bed. Secure it with end flaps or straps. Air tubes must be at foot end of
bed.
3. Connect air feed tubes to air outlet on the mattress and pump, ensuring that the tubes are not
"kinked" or twisted. Allow space for bed rails to drop freely.
4. Plug the pump into a suitable electrical power source and switch on. The switch indicator
should now illuminating.
5. Allow 20 minutes for mattress to inflate fully.
6. Cover with a sheet and tuck loosely.
7. Please use pressure adjustable knob to give maximum patient comfort.
Technical Specifications
Item: pM-2012 Plus Pump
Power supply: 220V / 50Hz
Air output: 5 liter/min.
Pressure range: 40 mmHg - 130 mmHg
Cycle time: 12 minutes
Power consumption: 7W
Fuse: T1A/250V
Cleaning Guidelines
• Pad can be cleaned by disinfectant solution.
• Do not immerse the pump in water.
• Do not heat or steam in autoclave.
Warranty for Pump and Mattress: 2 year
Pump
Item: pM-2012 Plus Mattress
Size: 280cm(L) x 90cm(W) x 6.5cm(H)
130 bubble cells
Without vents and with end flaps
Material: PVC (EN-71, low toxic)
Repair kit included