2.4

Unit book

We recommend to keep a unit book. All tests and essential works (e.g. repair work, modifications) must be documented
in this book.
2.5

Safety instructions

The effectiveness of the safety instructions with regard to personnel protection, handling the unit and handling the processed
product is to a large degree dependent on the behaviour of the staff working with the equipment.
Caution!
Prior to initial operation, carefully read through these Operating Instructions and observe the information
GB
in order to avoid errors and subsequent damage, particularly damage to health.
In addition to the instructions in this operating manual, comply with the national laws, regulations,
- 4 -
and directives in your country when setting up and operating the unit.
Hot surface!
The metal surface of the pressure pot may heat up considerably similar to a cooking pot.
Avoid touching the surface.
Caution!
Power cable and plug must be checked for damage prior to operation.
If any damage is found, the unit must not be connected to the mains.
Work on electronic equipment in the unit must be carried out by Heraeus Kulzer, Heraeus Kulzer service partners or qualified,
specialist personnel only and when the equipment is in a safe status (de-energised).
Only use authorised original spare parts and accessories. Avoid other parts as they hide unknown risks.
The operability and safety of the unit is only guaranteed if the necessary tests, maintenance and repair work is carried out
by Heraeus Kulzer, Heraeus Kulzer service partners or trained, specialist personnel.
For possible damage coming from a fault / malfunction of the unit due to improper repair, which was not carried out by
Heraeus Kulzer service partners or by personnel trained by us or in cases where original spare / accessory parts were not used
during a part replacement, Heraeus Kulzer GmbH is not liable.
Caution!
Be careful when filling and emptying the pot!
Risk of short circuit / electric shock owing to water which has penetrated into the unit!
Do not spill water! If required dry the housing with a cloth!
3

Use in accordance with specifications

The Palamat elite polymerization unit is a laboratory unit for polymerizing cold- and hot-curing resins for dental applications.
The functional design of the unit is specifically matched with processing methods of cold-and hot-curing resins.
The polymerization is performed at an operating pressure of 2 bars which is calibrated by an integrated pressure reduction
valve. A pressure gauge on the operating panel shows the nominal pressure after pressurization.
At a pressure of more than 2.8 bars within the pressure pot the safety valve opens automatically.
3.1

Working rules

Attention!
The following must be observed when working with the unit:
Attention!
Distilled, deionized water or other liquids may not be used!
The unit must not be used to heat up or prepare food.
Do not process or use readily flammable materials, liquids or gases.
Avoid the formation of explosive mixtures.
Do not use materials which may cause uncontrolled volume expansion in the pressure pot.
Use only clean tap water for processing.