Starting, fi nal test and users instruction
Install and connect the aspirator, select the position ON on
the switch, which will light up after one of the dental units
has started working. At this point aspiration will start.To
check if Turbo-Smart is working correctly, it is advisable to
carry out the dynamic tests (see fi g. D, page 123 and the
Turbo-Smart working diagram fi g. E, page 124). Users must
be instructed on the aspirator use and routine maintenance
of new, not used, and therefore not yet contaminated, ma-
chines. Show users how to follow the Turbo-Smart working
phases on the display, to interpret danger warnings and to
carry out routine maintenance using Puli-Jet plus new with
anti-scaling agent (A) - (by means of Pulse-Cleaner) (B) -
and disinfectant Antifoam Tablets (C).
Working
The aspirating unit (9) (through pipe 10) creates vacuum
inside the centrifugal separator (11). The fl uid coming from
the dental units enters the centrifugal separator (11) from
the collecting pipe (2). The centrifugal separator separates
air from liquids: air is exhausted outside through the pipe
(4) whereas liquids (in the version without amalgam sepa-
rator) are drained to the sewage through the waste pipe
connected to the tube-holder (6). The centrifugal separator
(11) is starting before the aspirating unit (9), this allows to
drain the liquids which might collect inside the centrifugal
separator before aspiration starts. Moreover, when the ma-
chine is switched off a timer (adjustable) is keeping the mo-
tor running for min.10'' - max. 120''.
Amalgam Separator ISO 18
On request, Turbo-Smart can be supplied with the amalgam
separator "Hydrocyclone ISO 18" having its separate manual.
B
C
2
6
4
A
11
9
10
31