4. SETTINGS
The scale allows the user to customize several settings including check weigh limits,
backlight operation and calibration.
4.1 Over and Under Limits
Press and hold SELECT, and while still pressing SELECT, press ZERO until [SET--HL]
is displayed.
Press ZERO to view the upper limit [000000].
Press SELECT ( ), TARE ( ), M+ ( ) to edit the digits of the upper limit.
Press ENTER (ZERO Key) to confirm the setting and display the lower limit
[000000].
Press SELECT ( ), TARE ( ), M+ ( ) to edit the lower limit.
Press ENTER (ZERO Key) to confirm the lower setting and display will show Run---X
where x is the alarm setting. In this case x=
0 = Turns off check weigh mode
1 = under / accept / over indicators are displayed, alarm beeps when weight is
outside accept range.
2 = under / accept / over indicators are displayed, alarm beeps when weight is
within accept range.
3 = under / accept / over indicators are displayed, alarm is off.
Press TARE & M+ to change the alarm to desired setting.
Press ENTER (ZERO key) to confirm setting and display [EndSEt]
When [EndSEt] is displayed, Press and hold SELECT, and while still pressing
SELECT, press ZERO key to return to weighing.
4.2 Backlight
Press and hold ZERO button until [bL---x] is displayed, where x=
0 = backlight is turned off
1 = backlight is on during weighing (15 seconds of no activity)
2 = backlight is always on
Press TARE to change the setting. Press ZERO to confirm the setting and return to
weighing.
1000 Series
EN-7