9. Electromagnetic Compatibility
Product has been tested for compliance with current regulatory
standards regarding its capacity to block EMI (electromagnetic
interference) from external sources.
Some procedures can help reduce electromagnetic
interferences:
• Use only Arjo cables and spare parts to avoid increased
emissions or decreased immunity which can compromise the
correct functioning of the equipment.
• Ensure that other devices in patient-monitoring and/or life-
support areas comply to accepted emissions standards.
Warning
Wireless communications equipment such as wireless computer
network devices, mobile phones, cordless telephones and their base
stations, walkie-talkies, etc. can affect this equipment and should be
kept at least 1.5m away from the equipment.
Intended Environment: Professional Healthcare Facility
Environment.
Exceptions: HF Surgical Equipment and the RF Shielded room
of an ME SYSTEM for magnetic resonance imaging.
Warning
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
54