WARRANTY
DeVilbiss provides warranty for this device . Use only DeVilbiss approved original spare parts for maintenance and repair to retain warranty . Using unapproved or non-original
spare parts would void warranty .
PROVIDER'S NOTES - Cleaning and Disinfection When There is a Patient Change
When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary) . The
next patient, user or third party should be protected by appropriate handling and preparation .
Therefore, when there is a patient change, people must be protected during the transport and handling of the device, and the device must be fully processed, i .e ., cleaned and
disinfected, by suitably trained personnel before reuse to protect the next patient. The complete processing may only be done by the manufacturer or by a qualified DeVilbiss
provider/service technician .
NOTE– If the following described complete processing of the concentrator by a qualified DeVilbiss provider/technician is not possible, the device must not be used by another
patient!
DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients.
NOTE–If preventive maintenance is due at this time, these procedures should be carried out in addition to the servicing procedures.
1.
Dispose of all accessory components that are not suitable for reuse, i.e., particularly the oxygen tubing, the nasal cannula/mask, oxygen outlet connector and humidifier
bottle .
2 .
The concentrator must be disconnected from the power supply for this step: Open the concentrator and remove all dust deposits inside the cabinet with an
appropriate vacuum cleaner .
3 .
Clean and disinfect the power cord and all internal and external parts with a suitable disinfecting agent, e .g ., Microbac Forte or Terralin® .
4 .
Check the cord, the plug on the back of the device, the power switch, the fuse holder and the indicator light for possible damage .
5 .
Replace all damaged or worn components .
6.
Replace the cabinet air filter on the back of the device.
7.
Check the oxygen concentration. If the device is within specifications, the extended life intake bacteria filter does not need to be replaced between patients. If the
concentration is not within specifications, the provider should refer to the service manual section on Troubleshooting.
NOTE– There is no portion of the gas pathways through the concentrator that could be contaminated with body fluids under normal conditions.
The device patient connection may unintentionally become contaminated with expired gases for a single fault condition i .e ., a hose internal to the device becomes
disconnected. This condition will cause no flow out of the device and/or an alarm condition. Should this occur refer to the Service Manual for additional instructions.
TROUBLESHOOTING
The following troubleshooting chart will help you analyze and correct minor oxygen concentrator malfunctions . If the suggested procedures do not help, switch to your reserve
oxygen system and call your DeVilbiss homecare provider . Do not attempt any other maintenance .
WARNING
To avoid electric shock hazard, do not remove the cabinet. The cabinet should only be removed by a qualified DeVilbiss homecare technician.
Minor Troubleshooting Chart
SYMPTOM
A . Unit does not operate . Power light is off when
the power switch is "On ." Audible alert is
pulsing and Service Required light is flashing.
B . Unit operates, the Power light is on when the
power switch is "On ." Red Service Required
light is illuminated . Audible alert may be
sounding .
C . Unit operates, the power light is on when
power switch is "on," audible low-frequency
vibration sound is detected .
D . Both the green Normal Oxygen and the
yellow Low Oxygen lights are either on or off .
E . Yellow Low Oxygen light is on or the yellow
Low Oxygen light is on, and the intermittent
audible signal is sounding .
SE-525K3
POSSIBLE CAUSE
1 . Power cord not properly inserted
into wall outlet .
2 . No power at wall outlet .
3 . Oxygen concentrator circuit breaker
activated .
1. Air filter is blocked.
2 . Exhaust is blocked .
3 . Blocked or defective cannula, face
mask or oxygen tubing .
4. Blocked or defective humidifier
bottle .
5 . Flow meter set too low .
1 . OSD malfunction .
1 . Flow meter is not properly set .
2. Air filter is blocked.
3 . Exhaust is blocked .
1 . Check power cord connection at the wall outlet . On 230 volt units, also check
the mains connection on the back of the unit .
2 . Check your home circuit breaker and reset if necessary . Use a different wall
outlet if the situation occurs again .
3 . Press the concentrator circuit breaker reset button located below the power
switch . Use a different wall outlet if the situation occurs again .
If the above remedies do not work, contact your DeVilbiss provider .
1. Check the air filter. If the filter is dirty, wash it following the cleaning instructions
on page 6 .
2 . Check the exhaust area and make sure there is nothing restricting the unit
exhaust .
3. Detach cannula or face mask. If proper flow is restored, clean or replace if
necessary. Disconnect the oxygen tubing at the oxygen outlet. If proper flow is
restored, check oxygen tubing for obstructions or kinks . Replace if necessary .
4. Detach the humidifier from the oxygen outlet. If proper flow is obtained, clean
or replace humidifier.
5. Set flow meter to prescribed flow rate.
If the above remedies do not work, contact your DeVilbiss provider .
1 . Turn your unit "Off ." Switch to your reserve oxygen system and contact your
DeVilbiss provider immediately .
1 . Contact your DeVilbiss provider .
1. Ensure the flow meter is properly set to the prescribed number. (The maximum
flow meter setting is 3 LPM when an oxygen bottle is being filled with oxygen
from the auxiliary port .)
2. Check the air filter. If the filter is dirty, wash it following the cleaning instructions
on page 6 .
3 . Check the exhaust area and make sure there is nothing restricting the unit
exhaust .
If the above remedies do not work, contact your DeVilbiss provider .
REMEDY
EN - 7