b Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3 V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM –
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the CMS50E Pulse
Oximeter.
The CMS50E Pulse Oximeter is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of the CMS50E Pulse Oximeter can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
CMS50E Pulse Oximeter as recommended below, according to the maximum
output power of the communications equipment.
Rated
Separation distance according to frequency of transmitter(m)
maximu
150 kHz to 80 MHz
m output
power of
d
transmitt
er
(W)
0.01
0.058
0.1
0.18
1
0.58
10
1.83
100
5.8
3
5 .
V
1
80 MHz to 800
MHz
P
3
5 .
d
E
1
0.035
0.11
0.35
1.10
3.5
800 MHz to 2.7 GHz
d
P
0.07
0.22
0.7
2.21
7
7
P
E
1

Publicité