H. Testee can not use enamel or other makeup.
I. Please clean and disinfect the device after operating according to the User
Manual (7.1).
6.3. Clinical restrictions
A. As the measure is taken on the basis of arteriole pulse, substantial
pulsating blood flow of subject is required. For a subject with weak pulse
due to shock, low ambient/body temperature, major bleeding, or use of
vascular contracting drug, the SpO2 waveform (PLETH) will decrease. In this
case, the measurement will be more sensitive to interference.
B. For those with a substantial amount of staining dilution drug (such as
methylene blue, indigo green and acid indigo blue), or carbon monoxide
hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin,
and some with icterus problem, the SpO2 determination by this monitor
may be inaccurate.
C. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine
may also be a major factor blamed for serious error of SpO2 measure.
D. As the SpO2 value serves as a reference value for judgement of anemic
anoxia and toxic anoxia, some patients with serious anemia may also report
good SpO2 measurement.
7. Maintain, Transportation and Storage
7.1 Cleaning and Disinfecting
When using alcohol wipes to disinfect the device, please air dry or clean it
with clean soft cloth.
7.2 Maintain
A. Please clean and disinfect the device before using according to the User
Manual (7.1).
B. Please recharge the battery when the screen shows
C. Recharge the battery soon after the over-discharge. The device should
be recharged every six months when it is no regular used. It can extend
the battery life following this guidance.
7.3 Transportation and Storage
A. The packed device can be transported by ordinary conveyance or
according to transport contract. The device can not be transported
mixed with toxic, harmful, corrosive material.
B. The best storage environment of the device is -40 ºC ~ 60 ºC ambient
temperature and not higher than 95% relative humidity.
EN
.