DC CURRENT MEASUREMENT
WARNING:
Ensure that the test leads are disconnected from the meter before making current measure-
ments.
1
Set the function switch to 1000 A or 400 A or 40 A range position. and make sure that the current
under test does not exceed the upper limit of the measuring range you have selected. The meter
automatically defaults to DC Current.
2
Press the [DC ZERO] button once, the ∆ sign will
appear on the LCD indicating Range is zero.
3
Press the trigger to open up the transformer jaws and
clamp around the single conductor under test.
4
Read the display.
Note:
During current measurements keep the transformer
jaws fully closed, otherwise this will affect the
accuracy of the measurement.
When large currents the transformer jaws may buzz.
This is not a fault and does not affect the accuracy
of the reading.