PROGRAMMING AND VISUALIZATION OF ON-OFF ppM FUNCTIONING
PROGRAMMING AND VISUALIZATION OF PROPORTIONAL ppM FUNCTIONING
Rev 4.0
PROGRAMMING OF THE FUNCTIONING
Push and quickly release CAL button and scroll the menu using + and – buttons
till the display shows "Function Type". Push CAL to enter the submenu and
then, by pressing + and – button, choice "Function On-Off" and confirm pressing
CAL.
The display will show the value of setpoint stored in the last programming. Modify
it by pressing + and – button and confirm the choice pressing CAL.
The display will show the direction of dosing Chorination or Dechlorination
stored in the last programming. Choice the direction by + or – button and confirm
pressing CAL.
Select Dechlor if you want the pump to inject for ppM values greater than
Setpoint.
Select Chlor if you want the pump to inject for values of ppM lower than Setpoint.
The display will show the % of maximum frequency (150pulses/minute) stored in
the last programming. Modify it by pressing + and – button and confirm the choice
pressing CAL. Push ESC button to esc the menu and go back in measure with
the On-Off functioning just programmed.
In the On-Off functioning if the pump is dosing the display shows "On" and the %
frequency programmed on the first row, if the pump doesn't dose the display
shows "Off" and the % frequency programmed.
On the second row the display writes the measure.
Range of proportionality = 1ppM
Push and quickly release CAL button and scroll the menu using + and – buttons
till the display shows "Function Type". Push CAL to enter the submenu and
then, by pressing + and – button, choice "Function Prop" and confirm pressing
CAL.
The display will show the value of setpoint stored in the last programming. Modify
it by pressing + and – button and confirm the choice pressing CAL.
The display will show the direction of dosing Chlorination or Dechlorination
stored in the last programming. Choice the direction by + or – button and confirm
pressing CAL.
Select Dechlor if you want the pump to inject for ppM values greater than
Setpoint.
Select Chlor if you want the pump to inject for values of ppM lower than Setpoint.
The display will show the % of maximum frequency (150pulses/minute) stored in
the last programming. Modify it by pressing + and – button and confirm the choice
pressing CAL. Push ESC button to esc the menu and go back in measure with
the Proportional functioning just programmed.
In the Proportional functioning the display shows on the first row "Prop" and the
actual % frequency of dosing that the pump is doing respect to the % frequency
programmed.
On the second row the display writes the measure.
Cod. 00.008.212
13