Autonomy: 1 year to 2 years.
–
Working temperature range: +10°C to + 40°C.
–
Storage temperature range: -10°C to + 60°C.
–
Relative humidity: 80% to 100%, non-condensing
(depending on the model).
–
EN 61326-1 compliant electromagnetic compatibility.
–
Protection degree (according to CEI/IEC 60529):
IP51 or IP54 for the models fitted with a rubber
bellow.
–
OPTO-RS232 and OPTO-USB data outputs .
–
Refresh rate: 8 times per second (display and
OPTO output).
Model with a resolution to 0,01 mm/.0005 in
Numerical interval: 0,01 mm / .0005 in
Line scale: ±250 μm.
Scale interval: 0,01 mm/.0005 in.
Highest numerical value displayed:±9999,99 mm / ±99.9995 in.
Model with a resolution to 0,001 mm/.00005 in
Numerical interval: 0,01 mm / .0005 in
Line scale: ±25 μm.
Scale interval: 0,001 mm/.00005 in.
Displayed highest numerical value:±999,999 mm/±9.99995 in.
Additional technical data
Models 200-300
Measuring span (mm/in)
Resolution (mm/in)
Max. perm. error MPE
Repeatability MPE
(mm)
E
(mm)
R
12,5/.5
0,01/.0005
0,02
0,01
12,5/.5
0,001/.00005
0,008
0,002
19