8.2 Wiring plan of R-M input

The R-M switch located on the 15-pin/f connector serves for controlling the measuring
instrument from a programmable device. This switch has the same function as the key
or the foot switch.
M
Like the output of the classification through relays, the R-M switch is provided in standard
with the TT80 and TT90 for connecting used programmable devices to Vcc or GND.

8.3 Possible connections (examples)

For these connections, you may either use an internal or external power supply (48 V, max.
30 mA), but never both together.
– Select the HOLD mode.
– Use the M switch or the R-M command to get the measured value.
– Release either of both switches to enable the classification function.
– Wait at least 300 ms before capturing a new value.
R-M command «Sink»
+5V
6
14
R-M
7
GND/R-M
GND
12
5 ∏ 48V
R-M
14
7
GND/R-M
20
PC354
PC354
R-M command «Source»
+5V
6
R-M
14
7
GND/R-M
GND
12
5 ∏ 48V
R-M
14
GND
7
GND/R-M
PC354
PC354