TROUBLESHOOTING

Problem
LCD does not light
System will not actuate
System will not pressurize
System will not pull vacuum
Inconsistent dispensing
Possible cause
No power input
Foot switch not plugged in or
improperly plugged in
Defective foot switch
Broken wire or loose connection
inside unit
Defective solenoid
Defective PC board
Insufficient air pressure
Air hoses not plugged in
Regulator defective
Vacuum setting is too low
Defective solenoid
Defective vacuum venturi
Defective vacuum needle valve
Air bubbles in adhesive
Dispense time is too low
Dispense needle started to clog
Correction
Check power cord connections
Turn on power
Check foot switch connection
Foot switch needs to be repaired or
replaced
Unplug power cord and disconnect air
supply. Remove cover and check for
broken wires or loose connections
Replace solenoid
Replace PC board
Increase air supply pressure
Check connection
Replace regulator
Increase vacuum setting
Replace solenoid valve
Replace venturi
Replace needle valve
Reduce vacuum setting
Increase dispensing time
Replace needle
invicon product insert
invicon Product insert
Precision Despenser 100
100 Precision dispenser