PC diagnostic tool
Test procedure
12
All manuals and user guides at all-guides.com
The test starts. The entire test run takes approx. 20 minutes.
The result, i.e. "Pass" or "Fail" appears at the end of the test
(see sections entitled "Test result: "Pass" or "Test result: "Fail").
3. On completion of the test, press any key to go on to the test dialog.
Tip: You can scroll to the individual test steps using the arrow keys.
The corresponding result is shown in front of each test step.
4. To quit the test, reboot the PC (see Rebooting the PC [ ➙ 14]).
Test result: Pass
No fault was found in the PC-specific hardware. A replacement of the PC
component or of the PC makes no sense.
1. Perform the separate test for the supply board.
2. Perform an image restore with the restore set for troubleshooting.
3. Check the service instructions for other possible fault sources.
Test result: Fail
A fault was found in the PC-specific hardware. A replacement of the PC
component or of the PC makes sense when indicated.
An image restore makes no sense.
1. Check to find out which components did not pass the test.
2. Carry out the corresponding steps described in the chapter on Trou-
bleshooting [ ➙ 14] .
3. Note the number of the failed test and the error code if available. At-
tach this information to the returned PC when replacing a PC.
Sirona Dental Systems GmbH
Diagnostic Tool CEREC AC
63 24 599 D3492
D3492.076.03.01.02 03.2011