PC diagnostic tool
Test procedure
10
All manuals and user guides at all-guides.com
1. Check the information in the system configuration to determine whe-
ther the system corresponds to the default settings. You can scroll up
and down using the arrow keys.
Tip: The test will start automatically in 3 minutes if you do not press
an arrow key first.
2. Press the "Esc" key on the keyboard.
The test starts. The entire test run takes approx. 20 minutes.
The result, i.e. "Pass" or "Fail" appears at the end of the test
(see sections entitled "Test result: "Pass" or "Test result: "Fail").
3. On completion of the test, press any key to go on to the test dialog.
Tip: You can scroll to the individual test steps using the arrow keys.
The corresponding result is shown in front of each test step.
4. To quit the test, reboot the PC (see Rebooting the PC [ ➙ 14]).
Sirona Dental Systems GmbH
Diagnostic Tool CEREC AC
63 24 599 D3492
D3492.076.03.01.02 03.2011