Sirona Dental Systems GmbH
Diagnostic Tool CEREC AC
63 24 599 D3492
D3492.076.03.01.02 03.2011
All manuals and user guides at all-guides.com
4.3
Camera cable
1. Connect the replacement camera cable to the Bluecam camera and
the camera tester.
2. Open the CEREC application and take an exposure with the camera.
No camera image is displayed
The Bluecam camera is defective.
Replace the Bluecam camera
(Bluecam rep. 62 36 884 or Blue-
cam replacement 62 36 694).
Camera test
Camera cable
A camera image is displayed
The camera cable was the cause of
the fault. Replace the camera cab-
le with the one attached to the CE-
REC AC diagnosis tool.
Tip: Replace the camera cable re-
moved from the CEREC AC dia-
gnosis tool with a new replacement
camera cable (replacement Blue-
cam camera cable 62 35 845).
7