Performance Verification
Performance verification ensures that the instrument performs within the specifications stated in the data sheet.
Some specifications are included in this document and were accurate at the time of publication, but the latest data
sheet from the Keysight website should be referred to see if there are any changes since this document was
published.
Recommended test equipment
The test equipments recommended for the performance verification and adjustment procedures are listed below. If
the exact instrument is not available, use the accuracy requirements shown to select substitute calibration
standards.
Type
Digital Multimeter
Current Monitor
Electronic Load
LAN / USB Controller
Oscilloscope
RMS Voltmeter
Differential Amplifier
Terminations
Variable voltage transformer or
AC source
Calibration resistors for small
current
Test considerations
– Ensure that the calibration ambient temperature is stable and between 20 °C and 30 °C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
65
Specification
Readout: 6 1/2 digits Basic DC Accuracy: 0.0035%
15 A (0.1 Ω), TC = 4 ppm/°C
150 V, 5 A minimum, with transient capability and a slew rate of
833 kA/s or better.
Sensitivity: 1 mV; Bandwidth Limit: 20 MHz Probe: 1:1 with RF
tip
True RMS; Bandwidth: 20 MHz Sensitivity: 100 μV
Bandwidth: 20 MHz
1 – 50 Ω BNC termination
2 – 50 Ω, 1/8 W termination resistors
Adjustable to highest rated input voltage range. Power: 500 VA
300 Ω (E36102B) 1 W minimum
2.5 kΩ (E36103B) 1 W minimum
9 kΩ (E36104B) 1 W minimum
20 kΩ (E36105B) 1 W minimum
51 kΩ (E36106B) 1 W minimum
Recommended model
Keysight 34401A or equivalent
Guildline 9230A-15R
Keysight N3300A mainframe, with
N330xA modules
PC with Keysight Connection Expert
loaded
Keysight Infiniium/6054A or equivalent
Rhode and Schwartz Model URE3 or
equivalent
LeCroy 1855A or equivalent
Keysight 6813B or equivalent
Keysight E36100B Series Operating and Service Guide

Publicité

loading