OUTPut Subsystem
OUTPut[:STATe] ON | 1 | OFF | 0
OUTPut[:STATe]?
Enables or disables the instrument's output.The query returns 0 (OFF) or 1 (ON). At *RST, the output state is off.
OUTPut:PROTection:CLEar
Clears the latch that disables the output due to an overvoltage or overcurrent condition. You must clear the
conditions that cause the fault before executing this command. You can then restore the output to the state that
existed before the fault condition occurred.
OUTPut:PON:STATe RST|RCL0|RCL1|RCL2|RCL3|RCL4|RCL5|RCL6|RCL7|RCL8|RCL9|
OUTPut:PON:STATe?
Specifies whether the instrument's power-on state is the *RST state (default) or the state stored in one of ten
memory locations.
53
Keysight E36100B Series Operating and Service Guide