Télécharger Imprimer la page

GE Druck ADTS 405 R/F Mk2 Guide D'installation page 12

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 95
8.
Operating Environment
Storage
temperatures
Operating
temperatures/
environment
AC Power supply
DC Power supply
(optional 405F)
Electrical safety
Operating
altitude
Ingress
protection
* Refer to local guidelines and regulations for power supply and electrical
connection recommendations.
[EN] English
WARNING
THIS EQUIPMENT IS NOT RATED FOR USE IN POTENTIALLY
EXPLOSIVE ATMOSPHERES.
ADTS 405F: -51°C to 70°C (-60° to 158°F)
ADTS 405R: -51°C to 70°C (-60° to 158°F)
ADTS Remote Terminal: -51°C to 70°C (-60° to 158°F)
ADTS 405F: -40°C to +55°C (-40° to 131°F)
ADTS 405R: -20°C to 50°C (-4° to 122°F)
ADTS Remote Terminal: -40°C to 50°C (-40° to 122°F)
Operating humidity: 5% to 95% RH (condensing)
Indoor & Outdoor use
*Universal input Mains Power Supply.
Class 1 equipment requiring a protective earth connection.
†The supplied mains/PSU cable protective earth/ground
conductor must be connected to the power supply protective
earth/ground system.
TXAH 250V HRC Time-delay (T) 250V a.c. rated fuses
(5A fuse)
Overvoltage category II, Pollution degree 3
24 to 30 V
†The supplied mains/PSU cable protective earth/ground
conductor must be connected to the power supply protective
earth/ground system.
T20A / T32A 250V DC fuses
Pollution degree 3
EN 61010-1, IEC 61010-1
Maximum 15,000ft (4572 metres)
ADTS 405F: MIL-PRF-28800 Class 2
ADTS 405R: IP23
ADTS Remote Terminal: IP65 (EN60529)
8
124M8686 Revision A

Publicité

loading