Chlorine HR T / M103
Determination of free chlorine HR with tablet
Select the method on the device.
EN
Fill 24 mm vial with 10 mL
sample.
Zero
Press the ZERO button.
Add DPD No. 1 HR tablet.
28
10 mL
Close vial(s).
Remove the vial from the
sample chamber.
Crush tablet(s) by rotating
slightly.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Empty vial except for a few
drops.
24 mm
Fill up vial with sample to
the 10 mL mark.
EN   Method Reference Book   1.0