Chlorine L / M101
Determination of free chlorine with liquid reagent
Select the method on the device.
EN
Fill 24 mm vial with 10 mL
sample.
Zero
Press the ZERO button.
Hold cuvettes vertically
and add equal drops by
pressing slowly.
20
10 mL
Close vial(s).
Remove the vial from the
sample chamber.
Add 6 drops DPD 1 Buffer
Solution to the sample
vial.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Empty vial.
6
Sample
Add 2 drops DPD
1 Reagent Solution to the
sample vial.
EN Method Reference Book 1.0
Sample
2
Sample