Télécharger Imprimer la page

Tintometer Lovibond MD 100 Manuel page 24

Masquer les pouces Voir aussi pour Lovibond MD 100:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 146
Chlorine L / M101
Chemical Method
DPD
Appendix
Interferences
EN
Persistant Interferences
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
Concentrations above 4 mg/L Chlorine, in the event of using fluid reagents, can lead
to results within the measuring range of up to 0 mg/L. In this case, the sample must
be diluted with chlorine-free water. 10 ml of the diluted sample should be mixed with
the reagent and the measurement taken again (plausibility test).
Interference
CrO
2-
4
MnO
2
Conformity
EN ISO 7393-2
determination of free, combined and total
a)
24
from / [mg/L]
0,01
0,01
EN   Method Reference Book   1.0

Publicité

loading

Ce manuel est également adapté pour:

Lovibond md 110Lovibond md 200