Preparation
1. Cleaning of vials:
As many household cleaners (e.g. dishwasher detergent) contain reducing
substances, this can lead to lower results with the determination of chlorine. To avoid
measurement errors, the glassware used should be free of chlorine consumption. To
achieve this, all glassware should be placed in a sodium hypochlorite solution (0.1 g/
L) for one hour and then rinsed thoroughly with deionised water.
2. For individual testing of free and total chlorine, the use of different sets of glassware
is recommended (EN ISO 7393-2, 5.3)
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents
therefore contain a buffer for the pH adjustment. Strong alkaline or acidic water
samples must therefore be adjusted between pH 6 and pH 7 before the analysis (use
0.5 mol/L sulphuric acid or 1 mol/L sodium hydroxide).
Notes
1. Evo tablets can be used as an alternative to the corresponding standard tablet (e.g.
DPD No.3 Evo instead of DPD No.3).
EN   Method Reference Book   1.0
Chlorine HR T / M103
EN
27