Settings
Available Key
Settings
Current After List
DC or List
Voltage After List
Resistance After List
Power After List
Pace
Dwl or Trg
Current Mode
Fix, Stp, or Lst
Voltage Mode
Resistance Mode
Power Mode
90
Specifies what happens when the input sequence completes.
Default is DC.
Action required:
Press Curr List, Volt List, Res List, or Pow List to toggle between DC and List.
DC (Return to DC
Returns to the DC value that was in effect before the input sequence
Value)
started.
List (Last List Value) Remains at the last list value.
Configures the pacing of the step.
Default is Dwl.
Action required:
Press Pace to toggle between Dwl and Trg.
Dwl(Dwell)
The next step immediately inputs when the dwell time has elapsed.
Trg (Trigger)
The next step immediately inputs when an external trigger is
received. If the step time completes before trigger occurs, the step
remains at the last list value while waiting for the trigger.
Sets the current/voltage/resistance/power mode. This determines what happens to the
input current/voltage/resistance/power when the system is initiated or triggered.
Default is Fix.
Action required:
Press Curr Mode, Volt Mode, Res Mode, or Pow Mode to toggle between Fix, Stp, and Lst.
Fix (Fixed)
Keeps the input at its immediate value.
Stp (Step)
Steps the input to the triggered level when a trigger occurs
Lst (List)
Causes the input to follow the list values when a trigger occurs.