nor get it in touch with sterile personnel .
- The use of the stimulator can interfere with the functioning of Pacemakers or
other stimulators implanted or anyhow connected to the patient . Therefore
don't use the stimulator without consulting a specialist of these devices .
- Mobile, radio frequency communication devices (including cables for external
antennas and antennas) must not be used closer than 30 cm (12 inches) to
any part of the device, including the cables specified by the manufacturer.
Otherwise, the performance of the device may be affected .
- Use of this device near or over (above or below) other devices is prohibited
as it may cause improper functioning . If this type of use is necessary, the
device in question and the others involved must be closely monitored to
verify their correct functioning .
- The simultaneous connection of the patient to a surgical high frequency
device can cause burns in correspondence to the electrodes of the stimulator,
or anyway damages to the stimulator itself .
- Don't use the stimulator in presence of flammable gases.
- Verify that the cables don't touch with other cables or directly with the patient .
- Don't use the stimulator in case of any malfunctioning, leakage of battery
fluids, accidental pouring of liquids on the stimulator.
- Don't use non-original accessories . The use of non-original accessories can
modify the emission of the electrical impulse, increase electromagnetic
emissions and/or decrease the electromagnetic immunity of the device and
consequently affect the safety features of the stimulator itself .
- Do not act near microwave and/or short waves therapy devices .
- High risk of cardiac fibrillation due to application of electrodes in the thoracic area.
- Check the compatibility between the device and its accessories and handle
with particular care .
- The stimulation should not be applied across or through the head, directly on
the eyes, covering the mouth, on the front of the neck (especially the carotid
sinus), or from electrodes placed on the chest and the upper back or crossing
over the heart .
- Do not touch the battery contacts and the patient at the same time .
- To reduce the risk of liquid leakage from the battery, remove the battery
from the device when not in use for a long time .
Cleaning and maintenance
- The stimulator must be cleaned by a soft cloth dampen with natural water;
it's possible to use also degreasing or sterilizing substances . The usage of
ethylic alcohol is NOT allowed . Don't use sprays nor immerse the stimulator
in any liquid .
- A spilling or a stagnation of liquid or humidity on the device can cause a
malfunction of the stimulator, therefore it is important to avoid any spills of
liquid on the stimulator . In case of accidental spilling mop carefully any
eventual residue .
- For the device disposal please refer to the instructions in paragraph 4 . 6 .