Display Indications - BEHA Unitest TELARIS SCHLEIFE Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Display Indications

7.0 Display Indications
All segments are illuminated
The instrument is ready to
carry out measurements.
Contact voltage limit 50V
Active measurement
Contact voltage limit 50V
Resistance measurement
RPE/ RN/ RL
RPE = 1.88 Ω
Contact voltage limit 50 V
Resistance measurement
RPE/ RN/ RL
RL exceeded (>19.99 W)
Contact voltage limit 50 V
Resistance measurement
IR/IPSC
IPSC = 88A
Contact voltage limit 25 V
Resistance measurement
RS/ IK
Contact voltage limit exceeded
Contact voltage limit 50 V
30
Resistance measurement
RS/ IK
ZL = 2,88 Ω
Contact voltage limit 50 V
Resistance measurement
Rs/ IK
RL excessive (>19,99 Ω)
Measurement is interrupted
Contact voltage limit 50 V
Resistance measurement
RS/ IK
ZL excessive (>1,999kΩ)
Contact voltage limit 50 V
Socket error
Nominal voltage tolerance ex-
ceeded
Contact voltage limit 50 V
Socket Error
Faulty socket wiring (during
contact with PE electrode)
Contact voltage limit 50 V
PE Error
Contact voltage limit 50 V
Contact voltage limit excee-
ded
Battery / temperature / me-
mory display
Battery nearly empty
Excessive internal tempera-
ture Memory data